| Station name: | Aragats Neutron Monitor (Արագած) |
|---|---|
| Station short name: | ARNM |
| Start date: | 2000-04-01 |
| End date: | None |
| Head organization: | Cosmic Ray Division (CRD) of A.I. Alikhanyan National Science Laboratory, 2 Alikhanyan Brothers St.,Yerevan 0036, Armenia |
| Principal investigator: | Information withheld according to EU GDPR, please contact us at questions@nmdb.eu |
| Contact person: | Information withheld according to EU GDPR, please contact us at questions@nmdb.eu |
| Latitude [°]: | 40.47 |
| Longitude [°]: | 44.18 |
| Altitude [m] asl: | 3200.0 |
| Geomagnetic cutoff rigidity [GV]: | 7.1 |
| Average count rate [s-1] during solar activity minimum: | 690.0 |
| Average count rate [s-1] during solar activity maximum: | 590.0 |
| Reference atmospheric pressure [mbar]: | 690.0 |
| Barometric coefficient [%/mbar]: | 0.688 |
| Information on detector housing: | Roof construction: metallic tilt, wood, 8 g/cm^2 |
| Type of counter tube: | CHM-15, transimpedance preamplifier. Dead time of preamp: ~400ns |
| Number of counter tubes: | 18 |
| Information on detector: | Moderator: Polyethylene, Reflector: Polyethylene, 3 sections X 6 counters |
| Atmospheric pressure sensor: | 1.5% max. error, 0 to 85 C, sensitivity: 45mV/kPa, temperature compensated -40 to 125 C |
| Information about clock: | NTP server sincroniced by GPS |